LYBSY-3000  0.05% three-phase power standard source

LYBSY-3000  0.05% three-phase power standard source

LYBSY-3000  0.05% three-phase power standard source


1.IntroductionLYBSY-3000  0.05% three-phase power standard source

LYBSY-3000 is the use of modern testing, DDS waveform synthesis, high-speed digital processor (DSP), complex programmable logic array (CPLD), large-scale integrated power amplifier, and the design of embedded computer system technology. It is suitable for electric energy meter (optional), AC and DC indicating instrument test and calibration, is ideal equipment for power system the power of product verification and calibration.

2. Main functions and featuresLYBSY-3000  0.05% three-phase power standard source

Domestic system, measurement and signal integration in a module, the product of high integration, low failure rate, small volume, light weight, fast response speed, high efficiency, high reliability, strong function, high output power, output standard source.

Is the original operating system, starting immediately display the test picture, without guidance program, fast response speed, high working efficiency.

The windows and buttons (with two functions), diversity operation, suitable for different people and habits, simple operation.

The contents of AC / DC standard source, direct test of various DC indicating instruments.

The automatic verification energymeter (optional) and indicator indicators.

The voltage, current, power, phase, frequency, harmonic with superior closed-loop output, setting a place, easy to use software to adjust.

Thevoltage, current, phase with rich common practical, simple operation, a little place, convenient use and high efficiency.

Equipped with digital rotary encoder adjustment ,easy to use,simple.

The outputvoltage, current and power are high precision, high stability standard source,  softwarecalibration.

Output standard harmonic 2~31 times, can be single or arbitrary superposition of many harmonic output.

Between the three-phase voltage, three-phase current, between each phase voltage and current can be arbitrary phase shift, so it can simulate all kinds of fault output power.

Equipped with the three-phase frequency independent phase, frequency conversion.

Equipped with multiple alarm and protection functions, it can detect the fault automatically and show the fault type and location. It is safe and reliable.

The preparation of interface and software, open interface protocol, user programmable control instruments.

Can support domestic similar product operation software.

3. Main technical indexLYBSY-3000  0.05% three-phase power standard source

AC analog output

AC voltage output

span

100V 57.735V 380V 220V

Adjustment range

0120%

Adjust fineness

0.002%

Accuracy class

0.05%

stability

0.01%/2min

Distortion

≤0.1%Non-capacitive load

Output load

30VA/ Channel

AC current output

span

5A2A1A20A

Adjustment range

0-120%

Adjust fineness

0.002%RG

accuracy

0.05%RG

stability

0.01%/2min

Distortion

≤0.1%Non-capacitive load

Output load

25VA/ Channel

Output power

Active power accuracy

0.05%RG

Reactive power accuracy

0.1%RG

stability

0.01%/2min

Phase output

Adjustment range

359.99°

Resolution

0.01°

accuracy

0.05°

Power factor

Adjustment range

-10+1

Resolution

0.0001

accuracy

0.05%

frequency

Adjustment range

45Hz65Hz

Resolution

0.001Hz

accuracy

0.002Hz

Three - phase voltage, current symmetry and phase symmetry

Voltage, current symmetry

0.02

Phase symmetry

0.05°

Voltage and current harmonic output

harmonic order

231 times

Harmonic content

039%

Harmonic phase

359.99°adjustable

accuracy

0.2%

DC output (Optional function)

DC voltage

Voltage range

75mV10V100V300V600V

Output range

0120%

accuracy      

0.05%

stability

0.01%/1min

Ripple content    

≤0.5%

Adjust fineness  

0.002%RG

Output load 

10VA

DC current

Current range     

20mA10mA1mA

Output range

0120%

accuracy      

0.05%

stability

0.01%/1min

Ripple content    

≤0.5%

DC measurement (Optional function)

DC current measurement

Measuring range

0~20mA

accuracy

0.05%

DC voltage measurement

Measuring range

0~10V

accuracy

0.05%

Amount of electrical energy pulse (Optional function)

Pulse output

Active pulse output

level 5V±5%Output current 100mA

Maximum pulse frequency5kHz

Pulse input

Active pulse intput

level 5V±5%input current 1mA

Maximum pulse frequency5kHz

Environmental conditions

Operating temperature

0℃~40

Relative humidity

≤85%

Storage conditions

-30℃~60

Other

Working power supply

AC220V±15%

volume

450×440×132mmweight:20kg

4、Instrument key instructionsLYBSY-3000  0.05% three-phase power standard source

Key

instruction

VRange

Voltage range switching

IRange

Current range switching

V/Y

Three  phase four  wire and three  phase three wire

 Switching

Complete wiring conversionThe display status bar must have a V or Y display.

SET

In this series, the key is undefined

Zero

All the output down to zero, and cut off the source output, the equivalent of the source off, mainly for wire

MENU

menu

MODE

In this series, the key is undefined

Enter

enter

XB

Harmonics keys for setting harmonics.

U

Set, display, adjust the voltage

I

Set, display, adjust the current

P

Set, measure, display, adjust active power

Q

Set, measure, display, adjust reactive power

Φ

Set, display, adjust the phase

F

Set, display, adjust the frequency

A

Phase sequence indicator key

B

Phase sequence indicator key

C

Phase sequence indicator key

The cursor is shifted to the left; Dc range selection

The cursor is shifted to the right; Dc range selection

-

 minus

1】~【9

number key

 . 

Decimal point

0%】~【120%

Common voltage and current test point, press this key will also output the percentage at the same time

0.0L】~【0.0C

Commonly used capacitive, inductive load test points

5、Encoder  instructionsLYBSY-3000  0.05% three-phase power standard source

key

instruction

The encoder turns right

1 The number rises when the cursor is under the digit

2 DC output range selection

The encoder turns left

1 The number drops when the cursor is under the digit

2 DC output range selection

Press the encoder down

And the Enter key [Enter] function the same

6、Wiring instructions

AC voltage output

Wiring method

Y type wiringthree phase four wireUa Ub Uc Un

LYBSY-3000  0.05% three-phase power standard source

V type wiringthree phase three wire) Ua Uc Un

AC current output

Wiring method

Y type wiringthree phase four wireIa Ib Ic

LYBSY-3000  0.05% three-phase power standard source


V type wiringthree phase three wire) Ia Ic Ib Wiring short circuit

DC voltage output

Wiring method

Connect the Uo+ and the Rs+ to the + terminal of the test device ;

Connect the Uo- and the Rs- to the - terminal of the test device ;

LYBSY-3000  0.05% three-phase power standard source


DC current output

Wiring method

I + terminal connects to the test device + terminal

I - terminal connects to the test device - terminal

LYBSY-3000  0.05% three-phase power standard source


Dc voltage current

Wiring method

Green: DC voltage input

Black: DC voltage, current lowRed: DC current input

LYBSY-3000  0.05% three-phase power standard source


7、Instrument interface operation instructions

7.1Menu selection interface

At any interface, press the [MENU] key pop-up menu to select the interface:

LYBSY-3000  0.05% three-phase power standard source

7.2AC output interface

After power on, enter the standard output interface, as follows:



V

0.000

0.000

0.000

DC

0.001

A

0.000

0.000

0.000

0.001

φU

0.000

0.000

0.000


φI

0.000

0.000

0.000

50.000Hz

W

0.000

0.000

0.000

0.000

Vr

0.000

0.000

0.000

0.000

VA

0.000

0.000

0.000

0.000

PF

0.000

0.000

0.000

0.000


100V   5A   Type YNo harmonicOpen loop

AC output interface operating instructions:

Gear selection of voltage and current

pressVRange

Switching voltage limit

pressIRange

Switching current limit

Quick output of voltage

pressdigit】【U】【Enter

Simultaneous rise of threephase voltage

pressdigit】【U】【A】【Enter

rise only Ua=digit

pressdigit】【U】【B】【Enter

rise only Ub=digit

pressdigit】【U】【C】【Enter

rise only Uc=digit

Quick output of current

pressdigit】【I】【Enter

Simultaneousrise of  threephase current

pressdigit】【I】【A】【Enter

rise only Ia=digit

pressdigit】【I】【B】【Enter

rise only Ib=digit

pressdigit】【I】【C】【Enter

rise only Ic=digit

Voltage and current angle setting

pressdigit】【Φ】【Enter

Set the threephase power factor angle =digit

Output frequency setting

pressdigit】【F】【Enter

Set the standard output frequency =digit

Coarse adjustment and finetuning of various parameters

pressU】【EnterEnter the display areaU=×××.××× V, Rotate the digital encoder, which will adjust the digit size of the cursor.Press 】【 key to move the cursor position, will realize the parameter of the coarse and fine tuning.

pressU】【Enter

Adjust the amplitude of the three-phase voltage simultaneously

pressU】【A】【Enter

Adjust the amplitude of A phase voltage

按【U】【B】【Enter

Adjust the amplitude of B phase voltage

pressU】【C】【Enter

Adjust the amplitude of C phase voltage

pressI】【A】【Enter

Adjust the amplitude of A phase current

pressI】【B】【Enter

Adjust the amplitude of B phase current

pressI】【C】【Enter

Adjust the amplitude of C phase current

pressΦ】【Enter

Adjust voltage and current angle

pressF】【Enter

Adjust output frequency

Turn off source output

pressZero

Turn off source output

Three phase four wire and threephase three wire conversion

pressV/Y

Three phase four wire and threephase three wire switching

switch interface

pressMENU

Menu selection interface






7.3DC output interface

                                                                          0.0000V

gear

DC voltage 600V

Gear selection


status

DC source is open, DC voltage output

DC output interface operating instructions

DC gear selection

PressVRange

Use the [←] [→] key to select the voltage range

PressIRange

Use   key to select the current range

DC voltage output

Pressdigit】【U】【Enter

output voltageU =digit

Press0%】~【120%Common voltage test point

The percentage voltage of the output gear

PressU】【Enter

Adjust voltage value by knob

DC current output

Pressdigit】【I】【Enter

Output current I =digit

Press0%】~【120%Common current test points

The percentage current of the output gear

PressI】【Enter

Adjust current value by knob

Turn off DC source

PressZero

DC output wiring

DC current wiring: the connection cable to the front panel, DC current output terminals, red terminal for the positive,

Black terminal is negative.

DC voltage wiring: DC voltage output using four-wire output, where UO +, UO- for the outputterminal.

7.4Harmonic display interface

Order

UA[%]

10+

20+

IA[%]

10+

20+

Total

0.000

THDU

0.000

0.000

THDI

0.000

1

100.00

0.000

0.000

100.00

0.000

0.000

2

0.000

0.000

0.000

0.000

0.000

0.000

3

0.000

0.000

0.000

0.000

0.000

0.000

4

0.000

0.000

0.000

0.000

0.000

0.000

5

0.000

0.000

0.000

0.000

0.000

0.000

6

0.000

0.000

0.000

0.000

0.000

0.000

7

0.000

0.000

0.000

0.000

0.000

0.000

8

0.000

0.000

0.000

0.000

0.000

0.000

9

0.000

0.000

0.000

0.000

0.000

0.000

10

0.000

0.000

0.000

0.000

0.000

0.000

A phase

F content

P add

Q delete











Harmonic display operation instructions

ABC phase switch

PressA】【B】【C

Switch the current display

Add \ remove harmonics

PressP】【Q

Enter interface

7.5Vector display interface


UA

0.00

UB

0.00

UC

0.00

IA

0.00

IB

0.00

IC

0.00

a

0.00

b

0.00

c

0.00

F customary law

P angle

Q 0~360





Vector display operation instructions

UA.UB.UC

Three phase voltage Angle

IA.IB.IC

Three phase current Angle

a.b.c

The angle between three phase voltage and current

7.6electric energy calibration interface

Electric energy error


running

Measured electrical energy


Active power

Voltage

0.000

0.000

0.000

constant

0.0

Current

0.000

0.000

0.000

Number of turns

0

Voltage phase

0.000

0.000

0.000

Low frequency output

Current phase

0.000

0.000

0.000

50.000Hz

Active power

0.000

0.000

0.000

0.000

Reactive power

0.000

0.000

0.000

0.000

Apparent power

0.000

0.000

0.000

0.000

Power factor

0.000

0.000

0.000

0.000

State

100V   5A   Type YNo harmonicOpen loop








Electric energy calibration method:

Press 【→】 key, the cursor enters the editing state,then press the number key on the panel.Press 【→】 key, the cursor enters the next edit box,When the parameters are all set up,press [Enter] to confirm and start the verification.

Active energy calibration and reactive energy calibration, switching method:

Press 【→】 to the active energy, press 【Enter】 key to switch active or reactive.

Press 【→】 to the constant, set the constant, press 【Enter】 to determine.Press 【→】 to the number of turns, set the number of turns, press 【Enter】 to determine.

Press 【→】 to stop the verification, press 【Enter】 to start the calibration,

Constant setting range

set range(1A-18000;2A-9000;5A-3600;20A-900).

Number of turns setting range

set range (1---9999999)Generally set about 5 laps, refresh soon, the number of laps set bigger.

Pulse port definition


LYBSY-3000  0.05% three-phase power standard source

1 pin :  signal  (green)

3 pin :  5V    (red)

5 pin :  ground    (black)


Wiring method of active electric energy meter


LYBSY-3000  0.05% three-phase power standard source

Wiring method of passive electric energy meter


LYBSY-3000  0.05% three-phase power standard source

7.7Parameter calibration interface

This interface is for calibration of instrument parameters. In order to ensure the accuracy of the instrument, this interface is not open to the user.

8、Attention

The instrument must have a good grounding in use

Check the temperature of 23 + 1 degrees.

To ensure the accuracy of the instrument, please preheat 30 minutes before use.

Attention should be paid to the proper limits of the objects to be measured.

instrument terminal output as standard source, the terminal can not access any other power supply.

tandard power output, you must first select the output voltage and phase, power loss

Out of the maximum 120% of the theoretical value.

In this equipment and other equipment shall disconnect all power devices connected before communication, however electrical connections cause damage to the equipment.

When DC output 5A 25A, please raise the current to 100%, preheat five to ten minutes.